Full Text Available
Note: Clicking the button above will open the full text document at the original institutional repository in a new window.
An anatomical and physiological background to the heart as a pump, followed by a description of the causes and results of cardiogenic shock is given. A survey of the achievements of other researchers in this field is included. The design of a system for controlling and monitoring the Intra-Aortic Ba...
| Main Author: | |
|---|---|
| Other Authors: | |
| Format: | Thesis |
| Language: | English |
| Published: |
Department of Electrical Engineering
2016
|
| Subjects: | |
| Tags: |
No Tags, Be the first to tag this record!
|
| _version_ | 1867613183750963200 |
|---|---|
| access_status_str | Open Access |
| author | Holmes à Court, Robert Leonard |
| author2 | Besseling, Johann L N |
| author_browse | Besseling, Johann L N Holmes à Court, Robert Leonard |
| author_facet | Besseling, Johann L N Holmes à Court, Robert Leonard |
| author_sort | Holmes à Court, Robert Leonard |
| collection | Thesis |
| description | An anatomical and physiological background to the heart as a pump, followed by a description of the causes and results of cardiogenic shock is given. A survey of the achievements of other researchers in this field is included. The design of a system for controlling and monitoring the Intra-Aortic Balloon Pump is described. A narrow band-pass filter is used to reduce noise and limit the input to the detector to the spectrum within which the R-wave of an ECG signal lies. A safety circuit to prevent the heart pumping against an inflated balloon is incorporated. History of the 8 secs preceding any instant is available. A description of the medical tests carried out so far is given. |
| format | Thesis |
| id | oai:open.uct.ac.za:11427/22178 |
| institution | University of Cape Town (South Africa) |
| language | eng |
| last_indexed | 2026-06-10T12:32:06.010Z |
| license_str | Not specified — see source repository |
| provenance_str_mv | Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository |
| publishDate | 2016 |
| publishDateRange | 2016 |
| publishDateSort | 2016 |
| publisher | Department of Electrical Engineering |
| publisherStr | Department of Electrical Engineering |
| record_format | dspace |
| source_str | UCTD — University of Cape Town Open Access Repository |
| spelling | oai:open.uct.ac.za:11427/22178 Control and monitoring of an intra-aortic balloon pump for cardiac assist Holmes à Court, Robert Leonard Besseling, Johann L N Electrical Engineering Medical instruments and apparatus An anatomical and physiological background to the heart as a pump, followed by a description of the causes and results of cardiogenic shock is given. A survey of the achievements of other researchers in this field is included. The design of a system for controlling and monitoring the Intra-Aortic Balloon Pump is described. A narrow band-pass filter is used to reduce noise and limit the input to the detector to the spectrum within which the R-wave of an ECG signal lies. A safety circuit to prevent the heart pumping against an inflated balloon is incorporated. History of the 8 secs preceding any instant is available. A description of the medical tests carried out so far is given. 2016-10-19T03:51:39Z 2016-10-19T03:51:39Z 1971 Master Thesis Masters MSc (Eng) http://hdl.handle.net/11427/22178 eng application/pdf Department of Electrical Engineering Faculty of Engineering and the Built Environment University of Cape Town |
| spellingShingle | Electrical Engineering Medical instruments and apparatus Holmes à Court, Robert Leonard Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| thesis_degree_str | Master's |
| title | Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| title_full | Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| title_fullStr | Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| title_full_unstemmed | Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| title_short | Control and monitoring of an intra-aortic balloon pump for cardiac assist |
| title_sort | control and monitoring of an intra aortic balloon pump for cardiac assist |
| topic | Electrical Engineering Medical instruments and apparatus |
| url | http://hdl.handle.net/11427/22178 |
| work_keys_str_mv | AT holmesacourtrobertleonard controlandmonitoringofanintraaorticballoonpumpforcardiacassist |