A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting

Background: Laboratory biomarkers are important adjuncts to clinical data in diagnosing neonatal sepsis. Available diagnostic tests often provide results 6 to 48 hours later. A bedside C-reactive protein (CRP) test may be able to exclude or diagnose sepsis within minutes. Objectives: The objectives...

Full description

Saved in:
Bibliographic Details
Main Author: Prince, Kim Didi
Other Authors: Joolay, Yaseen
Format: Thesis
Language:English
Published: Department of Paediatrics and Child Health 2017
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613261122240512
access_status_str Open Access
author Prince, Kim Didi
author2 Joolay, Yaseen
author_browse Joolay, Yaseen
Prince, Kim Didi
author_facet Joolay, Yaseen
Prince, Kim Didi
author_sort Prince, Kim Didi
collection Thesis
description Background: Laboratory biomarkers are important adjuncts to clinical data in diagnosing neonatal sepsis. Available diagnostic tests often provide results 6 to 48 hours later. A bedside C-reactive protein (CRP) test may be able to exclude or diagnose sepsis within minutes. Objectives: The objectives were to validate the Alere AfinionTM point of care test (POCT) CRP in a tertiary neonatal unit against the gold standard CRP assay in use by the National Health Laboratory service and to determine the difference in time to obtaining a result between the two systems. Methods: A prospective observational study was conducted between February 2015 and June 2015. Neonates who were clinically indicated to undergo CRP testing were simultaneously tested using the POCT and laboratory assays. The sensitivities, specificities and predictive values for the POCT, with the laboratory test as the reference test were determined. The time to results between the two tests was compared. Results: There were 139 measured CRP sample pairs from patients with suspected or proven neonatal sepsis. Using 10 mg/L as the cutoff value for both CRP tests, the sensitivity, specificity, positive predictive value and negative predictive value were 97.4%, 99%, 97.4% and 99% respectively. The area under the receiver operating characteristic curve was 0.99 (p<0.001). The time to point of care result was 4 minutes. Laboratory results were registered at a mean of 4.7 hours but only checked after a mean of 6.8 hours. Conclusions: The POCT CRP and laboratory CRP test have excellent correlation in neonates and may be a useful, quick, reliable method to rationalise antibiotic usage, reduce costs and allow for earlier patient discharge.
format Thesis
id oai:open.uct.ac.za:11427/22823
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:33:19.547Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2017
publishDateRange 2017
publishDateSort 2017
publisher Department of Paediatrics and Child Health
publisherStr Department of Paediatrics and Child Health
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/22823 A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting Prince, Kim Didi Joolay, Yaseen Paediatric Medicine Background: Laboratory biomarkers are important adjuncts to clinical data in diagnosing neonatal sepsis. Available diagnostic tests often provide results 6 to 48 hours later. A bedside C-reactive protein (CRP) test may be able to exclude or diagnose sepsis within minutes. Objectives: The objectives were to validate the Alere AfinionTM point of care test (POCT) CRP in a tertiary neonatal unit against the gold standard CRP assay in use by the National Health Laboratory service and to determine the difference in time to obtaining a result between the two systems. Methods: A prospective observational study was conducted between February 2015 and June 2015. Neonates who were clinically indicated to undergo CRP testing were simultaneously tested using the POCT and laboratory assays. The sensitivities, specificities and predictive values for the POCT, with the laboratory test as the reference test were determined. The time to results between the two tests was compared. Results: There were 139 measured CRP sample pairs from patients with suspected or proven neonatal sepsis. Using 10 mg/L as the cutoff value for both CRP tests, the sensitivity, specificity, positive predictive value and negative predictive value were 97.4%, 99%, 97.4% and 99% respectively. The area under the receiver operating characteristic curve was 0.99 (p<0.001). The time to point of care result was 4 minutes. Laboratory results were registered at a mean of 4.7 hours but only checked after a mean of 6.8 hours. Conclusions: The POCT CRP and laboratory CRP test have excellent correlation in neonates and may be a useful, quick, reliable method to rationalise antibiotic usage, reduce costs and allow for earlier patient discharge. 2017-01-19T12:24:35Z 2017-01-19T12:24:35Z 2016 Master Thesis Masters MMed http://hdl.handle.net/11427/22823 eng application/pdf Department of Paediatrics and Child Health Faculty of Health Sciences University of Cape Town
spellingShingle Paediatric Medicine
Prince, Kim Didi
A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
thesis_degree_str Master's
title A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
title_full A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
title_fullStr A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
title_full_unstemmed A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
title_short A comparison of standard C-reactive protein laboratory measurement to point of care C-reactive protein test in a neonatal intensive care unit setting
title_sort comparison of standard c reactive protein laboratory measurement to point of care c reactive protein test in a neonatal intensive care unit setting
topic Paediatric Medicine
url http://hdl.handle.net/11427/22823
work_keys_str_mv AT princekimdidi acomparisonofstandardcreactiveproteinlaboratorymeasurementtopointofcarecreactiveproteintestinaneonatalintensivecareunitsetting
AT princekimdidi comparisonofstandardcreactiveproteinlaboratorymeasurementtopointofcarecreactiveproteintestinaneonatalintensivecareunitsetting