The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index

Background: We compared the accuracy and precision of estimation of the systolic blood pressure (SBP) by disappearance of the pulse oximeter trace (DOT), with noninvasive blood pressure (NIBP) measurement, across the range of body mass index (BMI), during spinal anaesthesia (SA) for caesarean sectio...

Full description

Saved in:
Bibliographic Details
Main Author: Samuel, Theresa
Other Authors: Dyer, Robert A
Format: Thesis
Language:English
Published: Department of Anaesthesia 2017
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613274669842432
access_status_str Open Access
author Samuel, Theresa
author2 Dyer, Robert A
author_browse Dyer, Robert A
Samuel, Theresa
author_facet Dyer, Robert A
Samuel, Theresa
author_sort Samuel, Theresa
collection Thesis
description Background: We compared the accuracy and precision of estimation of the systolic blood pressure (SBP) by disappearance of the pulse oximeter trace (DOT), with noninvasive blood pressure (NIBP) measurement, across the range of body mass index (BMI), during spinal anaesthesia (SA) for caesarean section (CS). Methods: Three groups of 25 parturients were recruited, with BMI of < 30- (Group 1), 30-40- (Group 2), and > 40 kg/m2 (Group 3) respectively. SBP was measured using the DASH® 3000 monitor (GE Health Care, UK) NIBP monitor, placed on the same arm as the pulse oximeter probe. Estimations of SBP were done before- and 5 minutes after induction of SA, during cuff inflation and deflation. The times taken for the estimations and the actual NIBP measurements were noted. Bland and Altman analysis was performed and the correlation coefficient estimated. Results: Concerning the most clinically relevant estimation, namely SBP during inflation post-SA: For Groups 1, 2 and 3, r = 0.56, 0.74 and 0.91; bias = -0.4, -2.9 and 0.8 mmHg, and limits of agreement = -27.7 to 26.9, -27.7 to 21.9 and -15.9 to 17.5 mmHg respectively. The mean (SD) time required for estimation of the SBP during inflation post- SA was 7.5 (1.1) s, 11.8 (3.8) s and 16.8 (4.2) s in the Groups 1, 2 and 3 respectively. The mean (SD) time required for measurement of post-SA SBP during inflations in Groups 1, 2 and 3, was 30.3 (13.1) s, 41.3 (10.2) s and 49.8 (14.6) respectively. In the post-SA period, mean time saved by estimating SBP during inflation was approximately 28.5 seconds, compared with 9 seconds during deflation. The time saved in Groups 1, 2 and 3 was 22.8 (13.2) s, 30.0 (11.6) s and 33.0 (15.6) s respectively. In the BMI > 40 kg/m2 group, the percentage error is ±13% of the mean systolic blood pressure observed, and the absolute error is ±16 mmHg, compared to ±27 mmHg in the normal BMI group. Conclusion: Post-SA estimation of SBP during cuff inflation in morbidly obese patients is more accurate and precise than in the other BMI Groups. Time to estimation is shorter than measurement by a clinically relevant period (33 s). This should improve patient safety in morbidly obese parturients.
format Thesis
id oai:open.uct.ac.za:11427/23653
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:33:31.121Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2017
publishDateRange 2017
publishDateSort 2017
publisher Department of Anaesthesia
publisherStr Department of Anaesthesia
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/23653 The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index Samuel, Theresa Dyer, Robert A Anaesthesiology Background: We compared the accuracy and precision of estimation of the systolic blood pressure (SBP) by disappearance of the pulse oximeter trace (DOT), with noninvasive blood pressure (NIBP) measurement, across the range of body mass index (BMI), during spinal anaesthesia (SA) for caesarean section (CS). Methods: Three groups of 25 parturients were recruited, with BMI of < 30- (Group 1), 30-40- (Group 2), and > 40 kg/m2 (Group 3) respectively. SBP was measured using the DASH® 3000 monitor (GE Health Care, UK) NIBP monitor, placed on the same arm as the pulse oximeter probe. Estimations of SBP were done before- and 5 minutes after induction of SA, during cuff inflation and deflation. The times taken for the estimations and the actual NIBP measurements were noted. Bland and Altman analysis was performed and the correlation coefficient estimated. Results: Concerning the most clinically relevant estimation, namely SBP during inflation post-SA: For Groups 1, 2 and 3, r = 0.56, 0.74 and 0.91; bias = -0.4, -2.9 and 0.8 mmHg, and limits of agreement = -27.7 to 26.9, -27.7 to 21.9 and -15.9 to 17.5 mmHg respectively. The mean (SD) time required for estimation of the SBP during inflation post- SA was 7.5 (1.1) s, 11.8 (3.8) s and 16.8 (4.2) s in the Groups 1, 2 and 3 respectively. The mean (SD) time required for measurement of post-SA SBP during inflations in Groups 1, 2 and 3, was 30.3 (13.1) s, 41.3 (10.2) s and 49.8 (14.6) respectively. In the post-SA period, mean time saved by estimating SBP during inflation was approximately 28.5 seconds, compared with 9 seconds during deflation. The time saved in Groups 1, 2 and 3 was 22.8 (13.2) s, 30.0 (11.6) s and 33.0 (15.6) s respectively. In the BMI > 40 kg/m2 group, the percentage error is ±13% of the mean systolic blood pressure observed, and the absolute error is ±16 mmHg, compared to ±27 mmHg in the normal BMI group. Conclusion: Post-SA estimation of SBP during cuff inflation in morbidly obese patients is more accurate and precise than in the other BMI Groups. Time to estimation is shorter than measurement by a clinically relevant period (33 s). This should improve patient safety in morbidly obese parturients. 2017-01-27T14:19:21Z 2017-01-27T14:19:21Z 2016 Master Thesis Masters MMed http://hdl.handle.net/11427/23653 eng application/pdf Department of Anaesthesia Faculty of Health Sciences University of Cape Town
spellingShingle Anaesthesiology
Samuel, Theresa
The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
thesis_degree_str Master's
title The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
title_full The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
title_fullStr The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
title_full_unstemmed The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
title_short The utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section: the effect of body mass index
title_sort utility of occlusion of the pulse oximeter trace in the estimation of systolic blood pressure during spinal anaesthesia for caesarean section the effect of body mass index
topic Anaesthesiology
url http://hdl.handle.net/11427/23653
work_keys_str_mv AT samueltheresa theutilityofocclusionofthepulseoximetertraceintheestimationofsystolicbloodpressureduringspinalanaesthesiaforcaesareansectiontheeffectofbodymassindex
AT samueltheresa utilityofocclusionofthepulseoximetertraceintheestimationofsystolicbloodpressureduringspinalanaesthesiaforcaesareansectiontheeffectofbodymassindex