The design and evaluation of the P₀.₁ method of assessing ventilatory drive

Hypercapnia and/or hypoxia normally cause hyperventilation, The sensitivity to hypercapnia and hypoxia and the resultant hyperventilation is reduced in a patient suffering from damped respiratory centre activity. The P 0.1 method is a modification of the rebreathing technique where a patient rebreat...

Full description

Saved in:
Bibliographic Details
Main Author: Gajjar, Narendrakumar Chhotalal
Other Authors: Amoore, John
Format: Thesis
Language:English
Published: Division of Biomedical Engineering 2019
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613248431325184
access_status_str Open Access
author Gajjar, Narendrakumar Chhotalal
author2 Amoore, John
author_browse Amoore, John
Gajjar, Narendrakumar Chhotalal
author_facet Amoore, John
Gajjar, Narendrakumar Chhotalal
author_sort Gajjar, Narendrakumar Chhotalal
collection Thesis
description Hypercapnia and/or hypoxia normally cause hyperventilation, The sensitivity to hypercapnia and hypoxia and the resultant hyperventilation is reduced in a patient suffering from damped respiratory centre activity. The P 0.1 method is a modification of the rebreathing technique where a patient rebreathes from a bag prefilled with a mixture of gases of known concentrations. This modification is not sensitive to airflow obstruction as is the case with the ventilatory response of the rebreathing technique. The P 0.1 method has been described by several authors during the past 8 years. There is however the need to standardise the technique. This thesis is an attempt to meet this need. The P 0.1 method involves the occlusion of the inspiratory airway for a set time after the onset of inspiration of a particular breathing cycle. The pressure (P 0.1) at this time, generated by the isometric contraction of the respiratory muscles, is recorded together with the end tidal P C02 . The valve is occluded approximately 20 times during a rebreathing trial lasting for 4 minutes. The plot of P 0.1 versus P C02 is linear and its slope gives an indication of respiratory centre activity. A low slope indicates damped respiratory centre activity. The control electronics was designed using digital logic. The occlusion valve is closed passively during the expiratory phase and then actively held for a preselected time, most frequently for 1OOms, after the onset of inspiration, during the next breathing cycle. The active occlusion period is preselectable between 50ms and 300ms. The inspiratory pressure is recorded at this time, or 10 or 20ms prior to the opening of the valve, by a sample and hold circuit. The occlusion valve can be triggered manually via a push button switch on the front panel, or periodically for preselectable periods every 5 to 30 seconds or pseudo-randomly. Ventilatory phase, pressure preset and autotrigger mode indicators are included as operator aids. In a limited number of clinical trials the equipment worked satisfactorily. During a hypercapnic study the CO2 concentration in the bag progressively increases with a resultant increase in ventilation. The hypercapnic trials carried out yielded encouraging results. The method is simple, rapid and easily reproducible. The regression line plots obtained are linear with correlation coefficients better than those presented in the literature. The sensitivity of the ventilatory drive defined as the slope of the P 0.1 versus P C02 regression line for a group of 10 healthy adult males tested was 0.79 ± 0.47 cmH 2 o/mmHg. The device is in clinical use in the Respiratory Clinic at Groote Schuur Hospital. Further work needs to be done to investigate the full meaning of the results obtained and to what extent it can be used as a noninvasive diagnostic and screening technique for respiratory disorders. The P 0.1 method will help the clinician to assess non-invasively the degree of impairment of respiratory centre output in patients suspected of having a damped respiratory centre and who may also suffer pulmonary mechanical defects because the technique is independent of airflow.
format Thesis
id oai:open.uct.ac.za:11427/30551
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:33:07.122Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2019
publishDateRange 2019
publishDateSort 2019
publisher Division of Biomedical Engineering
publisherStr Division of Biomedical Engineering
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/30551 The design and evaluation of the P₀.₁ method of assessing ventilatory drive Gajjar, Narendrakumar Chhotalal Amoore, John Biomedical Engineering Ventilators Mechanical Hypercapnia and/or hypoxia normally cause hyperventilation, The sensitivity to hypercapnia and hypoxia and the resultant hyperventilation is reduced in a patient suffering from damped respiratory centre activity. The P 0.1 method is a modification of the rebreathing technique where a patient rebreathes from a bag prefilled with a mixture of gases of known concentrations. This modification is not sensitive to airflow obstruction as is the case with the ventilatory response of the rebreathing technique. The P 0.1 method has been described by several authors during the past 8 years. There is however the need to standardise the technique. This thesis is an attempt to meet this need. The P 0.1 method involves the occlusion of the inspiratory airway for a set time after the onset of inspiration of a particular breathing cycle. The pressure (P 0.1) at this time, generated by the isometric contraction of the respiratory muscles, is recorded together with the end tidal P C02 . The valve is occluded approximately 20 times during a rebreathing trial lasting for 4 minutes. The plot of P 0.1 versus P C02 is linear and its slope gives an indication of respiratory centre activity. A low slope indicates damped respiratory centre activity. The control electronics was designed using digital logic. The occlusion valve is closed passively during the expiratory phase and then actively held for a preselected time, most frequently for 1OOms, after the onset of inspiration, during the next breathing cycle. The active occlusion period is preselectable between 50ms and 300ms. The inspiratory pressure is recorded at this time, or 10 or 20ms prior to the opening of the valve, by a sample and hold circuit. The occlusion valve can be triggered manually via a push button switch on the front panel, or periodically for preselectable periods every 5 to 30 seconds or pseudo-randomly. Ventilatory phase, pressure preset and autotrigger mode indicators are included as operator aids. In a limited number of clinical trials the equipment worked satisfactorily. During a hypercapnic study the CO2 concentration in the bag progressively increases with a resultant increase in ventilation. The hypercapnic trials carried out yielded encouraging results. The method is simple, rapid and easily reproducible. The regression line plots obtained are linear with correlation coefficients better than those presented in the literature. The sensitivity of the ventilatory drive defined as the slope of the P 0.1 versus P C02 regression line for a group of 10 healthy adult males tested was 0.79 ± 0.47 cmH 2 o/mmHg. The device is in clinical use in the Respiratory Clinic at Groote Schuur Hospital. Further work needs to be done to investigate the full meaning of the results obtained and to what extent it can be used as a noninvasive diagnostic and screening technique for respiratory disorders. The P 0.1 method will help the clinician to assess non-invasively the degree of impairment of respiratory centre output in patients suspected of having a damped respiratory centre and who may also suffer pulmonary mechanical defects because the technique is independent of airflow. 2019-09-18T10:33:58Z 2019-09-18T10:33:58Z 1983 2019-09-18T10:07:10Z Master Thesis Masters http://hdl.handle.net/11427/30551 eng application/pdf Division of Biomedical Engineering Faculty of Health Sciences
spellingShingle Biomedical Engineering
Ventilators
Mechanical
Gajjar, Narendrakumar Chhotalal
The design and evaluation of the P₀.₁ method of assessing ventilatory drive
thesis_degree_str Master's
title The design and evaluation of the P₀.₁ method of assessing ventilatory drive
title_full The design and evaluation of the P₀.₁ method of assessing ventilatory drive
title_fullStr The design and evaluation of the P₀.₁ method of assessing ventilatory drive
title_full_unstemmed The design and evaluation of the P₀.₁ method of assessing ventilatory drive
title_short The design and evaluation of the P₀.₁ method of assessing ventilatory drive
title_sort design and evaluation of the p₀ ₁ method of assessing ventilatory drive
topic Biomedical Engineering
Ventilators
Mechanical
url http://hdl.handle.net/11427/30551
work_keys_str_mv AT gajjarnarendrakumarchhotalal thedesignandevaluationofthep01methodofassessingventilatorydrive
AT gajjarnarendrakumarchhotalal designandevaluationofthep01methodofassessingventilatorydrive