Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achievi...
| Main Author: | |
|---|---|
| Other Authors: | |
| Format: | Thesis |
| Language: | English |
| Published: |
Department of Paediatrics and Child Health
2021
|
| Subjects: | |
| Tags: |
No Tags, Be the first to tag this record!
|
| _version_ | 1867613234595364864 |
|---|---|
| access_status_str | Open Access |
| author | Ismail, Mugammad Taib |
| author2 | Horn, Alan |
| author_browse | Horn, Alan Ismail, Mugammad Taib |
| author_facet | Horn, Alan Ismail, Mugammad Taib |
| author_sort | Ismail, Mugammad Taib |
| collection | Thesis |
| description | Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achieving effective phototherapy. Objectives To describe and compare the effect of distance, angle and plastic barriers on three different LED lights of different design. Methods Comparisons were made using a Servolite LED light, a General Electric (GE) Lullaby and a Ningbo David LED phototherapy light. Measurements were done according to methods described by the International Electrotechnical Communission (IEC). The effective irradiated area was measured on a grid measuring 60 x 30 cm subdivided into 5 x 5 cm squares. Measurements were done for the following scenarios: light placed at the manufacturers' recommended distance, 20 cm closer, 20 cm further, at an angle, through clear plastic and through scuffed perspex. Results When the lights were placed closer to the irradiated surface than the manufacturers' recommendations, the maximum irradiance increased, but the median irradiance and uniformity ratio decreased. When the lights were angled at 45 the median irradiance was decreased. A decrease in the median irradiance was also seen when phototherapy lights passed through scuffed plastic and food grade plastic. Conclusion Our study demonstrated that placing LED lights closer than the manufacturers recommendations, the use of transparent barriers and the use of lights at an angle, compromised phototherapy irradiance and distribution. Only the GE light met IEC standards. |
| format | Thesis |
| id | oai:open.uct.ac.za:11427/33717 |
| institution | University of Cape Town (South Africa) |
| language | eng |
| last_indexed | 2026-06-10T12:32:54.720Z |
| license_str | Not specified — see source repository |
| provenance_str_mv | Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository |
| publishDate | 2021 |
| publishDateRange | 2021 |
| publishDateSort | 2021 |
| publisher | Department of Paediatrics and Child Health |
| publisherStr | Department of Paediatrics and Child Health |
| record_format | dspace |
| source_str | UCTD — University of Cape Town Open Access Repository |
| spelling | oai:open.uct.ac.za:11427/33717 The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices Ismail, Mugammad Taib Horn, Alan paediatrics and child health Background Effective phototherapy reduces neonatal jaundice and its complications. Irradiance increases as the distance of the light source decreases from a single phototherapy light. There are limited studies of the effect of distance and positional changes on different LED light designs on achieving effective phototherapy. Objectives To describe and compare the effect of distance, angle and plastic barriers on three different LED lights of different design. Methods Comparisons were made using a Servolite LED light, a General Electric (GE) Lullaby and a Ningbo David LED phototherapy light. Measurements were done according to methods described by the International Electrotechnical Communission (IEC). The effective irradiated area was measured on a grid measuring 60 x 30 cm subdivided into 5 x 5 cm squares. Measurements were done for the following scenarios: light placed at the manufacturers' recommended distance, 20 cm closer, 20 cm further, at an angle, through clear plastic and through scuffed perspex. Results When the lights were placed closer to the irradiated surface than the manufacturers' recommendations, the maximum irradiance increased, but the median irradiance and uniformity ratio decreased. When the lights were angled at 45 the median irradiance was decreased. A decrease in the median irradiance was also seen when phototherapy lights passed through scuffed plastic and food grade plastic. Conclusion Our study demonstrated that placing LED lights closer than the manufacturers recommendations, the use of transparent barriers and the use of lights at an angle, compromised phototherapy irradiance and distribution. Only the GE light met IEC standards. 2021-08-06T08:56:42Z 2021-08-06T08:56:42Z 2021 2021-08-06T08:55:50Z Master Thesis Masters MMed http://hdl.handle.net/11427/33717 eng application/pdf Department of Paediatrics and Child Health Faculty of Health Sciences |
| spellingShingle | paediatrics and child health Ismail, Mugammad Taib The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| thesis_degree_str | Master's |
| title | The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| title_full | The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| title_fullStr | The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| title_full_unstemmed | The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| title_short | The effect of device position and use of transparent covers on the irradiance distribution of LED phototherapy devices |
| title_sort | effect of device position and use of transparent covers on the irradiance distribution of led phototherapy devices |
| topic | paediatrics and child health |
| url | http://hdl.handle.net/11427/33717 |
| work_keys_str_mv | AT ismailmugammadtaib theeffectofdevicepositionanduseoftransparentcoversontheirradiancedistributionofledphototherapydevices AT ismailmugammadtaib effectofdevicepositionanduseoftransparentcoversontheirradiancedistributionofledphototherapydevices |