Full Text Available

Note: Clicking the button above will open the full text document at the original institutional repository in a new window.

Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm

Includes abstract.

Saved in:
Bibliographic Details
Main Author: Esmail, Mohammed
Other Authors: Tsoeu, Mohohlo Samuel
Format: Thesis
Language:English
Published: Department of Medicine 2014
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1867613158378569728
access_status_str Open Access
author Esmail, Mohammed
author2 Tsoeu, Mohohlo Samuel
author_browse Esmail, Mohammed
Tsoeu, Mohohlo Samuel
author_facet Tsoeu, Mohohlo Samuel
Esmail, Mohammed
author_sort Esmail, Mohammed
collection Thesis
description Includes abstract.
format Thesis
id oai:open.uct.ac.za:11427/3382
institution University of Cape Town (South Africa)
language eng
last_indexed 2026-06-10T12:31:41.113Z
license_str Not specified — see source repository
provenance_str_mv Harvested via OAI-PMH from UCTD — University of Cape Town Open Access Repository
publishDate 2014
publishDateRange 2014
publishDateSort 2014
publisher Department of Medicine
publisherStr Department of Medicine
record_format dspace
source_str UCTD — University of Cape Town Open Access Repository
spelling oai:open.uct.ac.za:11427/3382 Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm Esmail, Mohammed Tsoeu, Mohohlo Samuel John, Lester Medicine Includes abstract. Includes bibliographical references. This dissertation investigates methods for improving the image quality of a digital radiography system. The Lodox Statscan™ X-ray system provides a full-body scanned image for initial diagnosis. The system is driven by a permanent magnet linear motor (PMLM) controlled by a cascaded proportional-proportional integral controller (P-PI). Transient errors in the trapezoidal motion profile of the scanning C-arm may cause mismatches between the detector and the collimated beams from the X-ray source. This results in a partial degradation of image quality. The Statscan™ X-ray system was investigated and the following constraints were identified: The scanning time is limited to 13 seconds and the maximum scan length is limited to 1.8 m. Since it was not possible to obtain the Lodox Statscan™ dynamics model, because of the similarity, a characteristic model was then developed using a DC motor in order to investigate the control dynamics. It is not advisable for a designer to manipulate the controllers on commercial machines except for changing the parameters for tuning. Therefore, a P-PI controller and a proportional-integral sliding mode controller (P-ISMC) as well as a Boundary Layer variant (P-ISMC+BD) were designed for fair comparison purposes. Root locus, Bode diagrams and integral sliding mode control methods, respectively, were used to design the P-PI and P-ISMC controller groups. Each controller consists of an inner loop and an outer position loop. Proportional integral (PI) and integral sliding mode controllers (ISMC) were used for the inner loop. The two inner loop controllers were tuned, and then tested, before cascading them with the outer position loop. The simulation and experiments were conducted to compare each controller’s performance on step set-point tracking, trapezoidal motion profile tracking, the time transient’s specifications and robustness against disturbances. In order to test image quality, 27 distance profiles were generated from P-PI, P-ISMC and P-ISMC+BD. In addition, four images captured by the Statscan™ were also selected. A time delay integrator (TDI) simulator was used on the distance profiles and the four images to generate 108 distorted image profiles. 2014-07-29T09:03:24Z 2014-07-29T09:03:24Z 2012 Master Thesis Masters MSc http://hdl.handle.net/11427/3382 eng application/pdf Department of Medicine Faculty of Health Sciences University of Cape Town
spellingShingle Medicine
Esmail, Mohammed
Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
thesis_degree_str Master's
title Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
title_full Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
title_fullStr Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
title_full_unstemmed Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
title_short Towards improving the Statscan X-ray image quality through sliding mode control of the C-arm
title_sort towards improving the statscan x ray image quality through sliding mode control of the c arm
topic Medicine
url http://hdl.handle.net/11427/3382
work_keys_str_mv AT esmailmohammed towardsimprovingthestatscanxrayimagequalitythroughslidingmodecontrolofthecarm