Full Text Available

Note: Clicking the button above will open the full text document at the original institutional repository in a new window.

Effects of the Lubo cervical collar on airway patency in awake adults – A magnetic resonance imaging study

Introduction: Intended for use by prehospital first responders, the Lubo TM cervical collar is an adjustable, radiolucent, single-use device that incorporates a mechanical jaw thrust mechanism. The combination enables non-invasive airway management in cases of trauma where cervical motion restrictio...

Full description

Saved in:
Bibliographic Details
Main Author: Jaga, Rudhir
Other Authors: Hofmeyr, Ross
Format: Thesis
Language:English
Published: Department of Anaesthesia and Perioperative Medicine 2023
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Introduction: Intended for use by prehospital first responders, the Lubo TM cervical collar is an adjustable, radiolucent, single-use device that incorporates a mechanical jaw thrust mechanism. The combination enables non-invasive airway management in cases of trauma where cervical motion restriction is necessary. The potential benefits include use as an airway adjuvant maintaining upper airway patency, reducing provider task loading. The limited research on the device efficacy and safety requires further investigation. Methods: A randomized , crossover, interventional study was performed to compare mean differences in airway patency at the level of the uvula, epiglottis, tongue and soft palate with and without the Lubo collar in awake volunteers using magnetic resonance imaging (MRI). Fourteen participants each underwent two MRI scans of the upper airway: A control scan with no Lubo collar, and an intervention scan with the Lubo collar applied and jaw thrust mechanism activated. Two independent radiologists measured anterior-posterior diameter of the airway at four anatomical levels on the resulting MRI images. Results: There was no significant difference in mean airway diameter between the control and intervention measurements at any level. Mean (SD; 95% CI: p-value) differences were 0.9 mm (-2.38; 2.3 to 0.5; p= 0.17) at the epiglottis, 0.5 mm (1.6; -0.5 to 1.4; p= 0.29) at the soft palate, 0.2 mm (2.86; -1.4 to 1.9; p = 0.78) at the tongue, 0.4 mm (4.04; -1.9 to 2.7; p = 0.72) at the uvula. Conclusion: The Lubo TM airway collar did not show a significant change in upper airway patency at four anatomical levels measured in awake adult participants. Further research is required to investigate its clinical use in patients that are unable to maintain upper airway tone. Groups of interest would include trauma, obstructive sleep apnoea, obesity and patients under general anaesthesia.