Full Text Available

Note: Clicking the button above will open the full text document at the original institutional repository in a new window.

Comparative evaluation of haemoglobin estimation amongst pregnant women in Ibadan: Hemocue – B Haemoglobin analyzer versus haemiglobincyanide (Standard) method as the gold standard

This was a comparative crossectional study that was conducted among 557 antenatal women at the University College Hospital, Ibadan, Nigeria between 1st April and 30th May 2006. The study compared the accuracy of haemoglobin estimation using an automated HemoCue B analyzer kit with a standard laborat...

Full description

Saved in:
Bibliographic Details
Format: Article
Published: 2008-08
Subjects:
Tags: Add Tag
No Tags, Be the first to tag this record!